Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.


Note
iconfalse
titleOverview

The PlasmaPro 100 Cobra ICP utilises high-density plasma to achieve high-rate etching. The process modules offer excellent uniformity, high-throughput, high-precision, and low-damage processes for wafer sizes up to 150 mm. 



Page properties
idequipment-properties


LMACS NameICPRIE (Cobra Reactive Etch)
Confluence Labeloxford-plasmapro-100-cobra
Process Area

Equipment process area
EquipmentAreaPlasma Etch

ModelPlasmaPro 100 Cobra
VendorOxford Instruments
Team



Show If
groupnf-staff


Serial Number220145
UofA Asset Tag Number605426
Service Contract

Status
colourRed
titleNO
. Warranty expires  

Cryo Table cooling Rigaku LN60




Show If
groupnf-staff

Contacts

Info

While there are many individuals within Oxford who we have contacted directly over the years, the names / numbers listed here are our official points of contact.



NameCompanyEmailPhone
Technical SupportCustomer SupportOIPTccc.oia@oxinst.com1-800-447-4717
Sales ContactDemetrius ChrysostomouOIPTdemetrius.chrysostomou@oxinst.com1-978-435-5734
Process SupportOIPT Plasma SupportOIPTptsupport@oxinst.com1-978-369-9933
E-store (Parts & Service)https://estore.oxinst.us/products/plasma-technology/




Equipment Configuration

Info
titleConfiguration Schedule

Please refer to the Oxford Cobra ICPRIE tool page on LMACS for the liner installation and 100 mm / 150 mm clamp change schedule.  By default, the Cobra is in 100 mm mode without liners.  Conversions to 150 mm mode can be done by LMACS request; please contact nanoFAB staff for more information.  

N.B.  Sputter-based etches (e.g., Au, Pt, Cu) are only allowed in the tool when liners are installed.  Users must book the tool using ICPRIE (Cobra Sputter Etch) in LMACS and only during the times when liners are installed.  The total volume of metal being etched may not exceed 0.04 mm³ (4e7 µm³) between cleaning cycles.  This corresponds to etching through a 100 nm thick film with an exposed area of 4 cm² (e.g., ~5% of a 100 mm wafer).

Different pre-process and post-process chamber cleaning recipes are required to be run depending on the tool configuration and the materials being processed.  Please contact a nanoFAB staff member if you are unsure of what the proper cleaning schedule for your process should be.


System Features

Wafer Handlingpieces (mounted)100 mm150 mm
Gas Species (Inert)O2CHF3SF6Ar
Gas Species (Toxic)Cl2HBrBCl3
ICP RF Power3 kW


Chuck RF Power300 W


Chuck-150 ºC (min)300 ºC (max)mechanical clampHBC


Documents

Operating Procedure
Hazard Assessment
Process Information


Show If
groupnf-staff

Staff Documents

Operator Manual

Google Drive Live Link
urlhttps://drive.google.com/a/ualberta.ca/file/d/0B800HgP1RkZLM1BIWTBXQXhqdzA/view?usp=drivesdk

Service Manual

Google Drive Live Link
urlhttps://drive.google.com/a/ualberta.ca/file/d/0B800HgP1RkZLazlaSkJKbHpjLTA/view?usp=drivesdk

Installation Guide

Google Drive Live Link
urlhttps://drive.google.com/a/ualberta.ca/file/d/0B800HgP1RkZLODFrTjZWd2dTMUk/view?usp=drivesdk

System Configuration

Google Drive Live Link
urlhttps://drive.google.com/a/ualberta.ca/file/d/0B800HgP1RkZLem1aWXdtTXgtbFU/view?usp=drivesdk

Process Gas Information

Google Drive Live Link
urlhttps://drive.google.com/file/d/0B800HgP1RkZLcFRhdnBILTFTNm11SVRVYzN0aU51bGlhR1Mw/view?usp=drivesdk

Drive Folder

Embedded Google Drive Folder
sort-modified
urlhttps://drive.google.com/drive/folders/1-rsbe30HqKiPa4VHadApT9jd5Ag6AZAa


Related Documents

Content by Label
cqllabel = "oxford-cobra-icp"